A method to minimize condenser lens-induced hysteresis effects in a JEOL JEM-3200FSC microscope to enable stable cryoEM low-dose operations. Gonen Lab
de la Cruz MJ, Martynowycz M, Hattne J, Shi D, Gonen T
janelia7_blocks-janelia7_biblio_abstract | block
Abstract
Low dose imaging procedures are key for a successful cryoEM experiment (whether by electron cryotomography, single particle analysis, electron crystallography, or MicroED). We present a method to minimize magnetic hysteresis of the condenser lens system in the JEOL JEM-3200FSC transmission electron microscope (TEM) in order to maintain a stable optical axis for the beam path of low-dose imaging. The simple procedure involves independent voltage ramping of the CL1 and CL2 lenses immediately before switching to the focusing and exposure beam settings for data collection.